
From: Mike Malin (malin@msss.com)
Subject: NEAR Calibration Update #4
Date: 10-15 January 1995
As noted in previous reports, I determined the mean and standard deviation for the 135 X 165 area. I then computed the averages of the means and standard deviations for all images taken during each test. Dark current signal vs. temperature is plotted in Figure 1 and dark current vs. shutter (exposure duration) is plotted in Figure 2.Dark Current at -30 deg C
I'm not sure why dark current seems to increase with decreasing temperature over these small Delta-T's, but this was also observed at -20 deg C. It may be a function of when or when T is being measured. As for the dark current accumulation rate, it's clear that for one set of points the change in temperature over the course of the measurements masked the steady-state accumulation rate, but the second set seems to give a reasonable result. From these figures, it appears that the average dark current signal for NEAR MSI images is about 40 DN, and the dark current accumulation rate at -30 deg C is about 1.3 DN/sec (which is about 1/2 to 1/3 of what was seen at -20 deg C).

Figure 1: Dark Current vs. Temperature at -30 deg C

Figure 2: Dark Current Accumulation Rate at -30 deg C
A more extensive (or at least complete) set of measurements were made of the transmission of the neutral density filters, at three specific spectral bandpass/shutter speed combinations: 500 nm @ 917 ms, 700 nm @ 70 ms, and 1000 nm @ 917 ms. The results of the ND Filter Transmission test are shown in Figure 3. Again, the characteristic curve is an exponential. The same discrepancies noted in Report 2 are seen here: the 700 nm/70 ms and 1000 nm/917 ms exposures are nearly the same DNs despite the difference in exposure time, and the relationships have slightly different slopes and intercepts. Again, I suspect that the 700 nm and 1000 nm filters have different widths. Note that in this example, I have included two fits to the data, one including and one excluding the ND 2.5 values. It appears to me that we will need a significantly brighter light source if we are indeed going to calibrate the 1.5, 2.0, and 2.5 ND filters.Neutral Density Filter Transmission at -30 deg C

Figure 3: ND Filter Transmission Test at -30 deg C
For completeness, I'll include two analyses I did of the exposure response of the CCD at -30 deg. The first two graphs (Figures 4 and 5) show the exposure response (linearity) and the standard deviation vs. mean brightness. Figure 4 shows the relationship between exposure duration and output DN, after correcting for dark current. With a readout rate of 1 ms, it is clear that the shift away from linearity occurs between 10 and 20 ms exposure.Exposure Response of CCD at -30 deg C

Figure 4: Mean Brightness versus Shutter Speed/Exposure Duration
Figure 5 shows the two, well-defined branches of the noise relationships: an essentially constant "read noise" component and power-law counting noise component. For reasons I'm investigating, however, this is not the square-root relationship it should be. [NB 1/15/95 - This and other photon-transfer curves are suspect and should be ignored for now. When I get this straightened out, look for a separate report.]

Figure 5: Standard Deviation of Brightness versus Mean Brightness
I had a hell of a time with this data set. There are no less than 7 ambiguous image pairs, summarized in the following table:Raw Spectral Response (i.e., No Lamp Correction) at -30 deg
Ambiguous Images at -30 deg C
File ID Filter Shutter ND filt Temp Mean StDev fumw301.355 500 100 0 -29.4 40.29 6.62 fumw302.355 500 100 0 -29.4 220.74 14.12 fumw341.355 500 500 1 -29.4 136.07 8.89 fumw342.355 500 500 1 -29.4 217.06 13.34 fumw541.355 600 500 1 -29.4 613.55 35.21 fumw542.355 600 500 1 -29.5 1089.66 62.62 fumw941.355 800 500 1 -29.4 819.65 37.18 fumw942.355 800 500 1 -29.5 1467.03 67.74 fumwb42.355* 900 500 1 -29.5 1091.54 42.46 fumwb11.355* 900 500 1 -29.4 613.99 24.21 fumwd41.355 1000 500 1 -29.4 203.27 9.29 fumwd42.355 1000 500 1 -29.5 339.66 14.10 fumwe41.355 1050 500 1 -29.4 85.14 6.80 fumwe42.355 1050 500 1 -29.5 121.80 6.76
Of these image pairs, the ID numbers of the ones marked with asterisks (*) decode differently than the label information: fumwb42.355 decodes to (900 nm, ND 1) while fumwb11.355 decodes to (900 nm, ND 0.1), although both have labels that say (900 nm, ND 1, 500 ms). The rest of these pairs have identical ID numbers (as far as image type is encoded in the ID number), identical labels, but difference means and standard deviations. Owing to these abiguities, it wasn't possible to put together any reasonable spectral curves for the -30 deg C measurements.
I computed the dark current signal and dark current accumulation rate as described in previous reports. Dark current signal vs. temperature is plotted in Figure 6 and dark current vs. shutter (exposure duration) is plotted in Figure 7.Dark Current at -40 deg C
Again, dark current seems to increase with decreasing temperature over these small Delta-T's; again, this must be due to some experimental setup relationship, since dark current shouldn't increase with lower temperature. Since I am already suspicious of the 917 ms values since they were taken at lower temperature but give a higher dark current signal, I believe the 100 and 500 ms measurements are more self-consistent. From these figures, it apparent that the average dark current signal for NEAR MSI images is again about 40 DN (~43 DN), and the dark current accumulation rate at -40 deg C is about 0.75 DN/sec (which is about 1/2 of what was seen at -30 deg C, and 1/4 of what was seen at -20 deg C).

Figure 6: Dark Current vs. Temperature at -40 deg C

Figure 7: Dark Current Accumulation Rate at -40 deg C
The results of the ND Filter Transmission test are shown in Figure 8. This figure basically shows the same results as seen at -20 and -30 deg C: the detector produces an exponential relationship between the neutral density filters. The 100 ms exposure at 700 nm produces a brightness only about 50% greater than the 917 ms exposure at 1000 nm because the 700 nm filter's full-width at half-maximum (FWHM) is 37.4 nm while the FWHM for the 1000 nm filter is only 9.2 nm.Neutral Density Filter Transmission

Figure 8: ND Filter Transmission Test Results
Quantum efficiencies and exposure response can be determined "graphically" by use of the "photon transfer curve" (Janesick, Klaasen, and Elliott, Opt. Eng. v. 26, no. 10, 972-980, 1987). This curve plots noise against signal, and absolute quantum measurements can be made if appropriate, calibrated light sources are viewed at appropriate wavelengths. [NB 1/15/95 - There appears to be a discrepancy between this and other papers published by the same author(s), where discussing the value by which to divide the standard deviation: other articles say to divide by sqrt(2) rather than by 2. I'm looking into this and will report back in a separate report discussing the photon-transfer curve.]Exposure Response of CCD
Figure 9 shows the relationship between exposure duration and output DN, after correcting for dark current. I've plotted this as a log-log plot to show the loss of linearity at low shutter speeds (low illumination in these examples), since when the exposure rate and readout rate are comparable, there are substantial variations in exposure across the image. With a readout rate of 1 ms, it is clear that the shift away from linearity occurs between 10 and 20 ms exposure.

Figure 9: Mean Brightness versus Shutter Speed/Exposure Duration
I next plot the standard deviation against the mean for each image acquired (Figure 10). These data clearly show the two, well-defined branches of the noise relationships: an essentially constant "read noise" component and power-law counting noise component. Note, however, that in this representation the counting statistics limit curve is not a square-root--this is because I haven't (couldn't) extract fixed-pattern noise. [NB 1/15/95 - actually, I don't know why this curve isn't correct. This and other photon-transfer curves are suspect and should be ignored for now. When I get this straightened out, look for a separate report.]

Figure 10: Standard Deviation of Brightness versus Mean Brightness
The final set of data acquired at -40 deg addressed spectral response (Figure 11). I haven't created lamp- and filter-corrections to these curves...I think people who've done this before should take a try at it (Jim, Mark, Lucy, Joe?). Here is a table of filter locations and full-widths at half-maxima (Scott gave these to me, and said the filters were more or less gaussian in shape):Raw Spectral Response (i.e., No Lamp Correction)
Spectral Filter Centers and Bandwidths
Band Center (nm) Band Width (nm)
408.8 37.1
506.7 33.8
601.1 35.9
704.7 37.4
801.3 9.7
902.3 10.5
1003.0 9.2
1052.3 11.4
(see Report #2 for the probably spectral shape of the lamp's light)

Figure 11: Raw Spectral Response at -40 deg C

Figure 12: Photon Transfer Curve from -30 deg minus -20 deg C Images